    A novel normalization technique for multispectral acquisition (MSA) in positron emission tomography (PET) imaging is described. MSA data are affected by spectral nonuniformity due to variations in detector characteristics. As with conventional methods, the normalization must preserve counts as it reduces nonuniformity but spectral shapes must be restored and symmetry distortions between mirror windows must be compensated. The proposed method achieves these goals by averaging each window pair over all detectors and by balancing mirror windows. The method was tested with a variety of spectral distortions and was found to be successful in restoring data uniformity. It is concluded that MSA PET imaging is possible with relatively nonuniform detector response, but with stable detector characteristics are essential
  • Keywords
    computerised tomography; data acquisition; radioisotope scanning and imaging; PET imaging; detector characteristics variations; medical diagnostic imaging; multispectral acquisition; normalization technique; nuclear medicine; positron emission tomography; spectral distortions; spectral nonuniformity; spectral shapes; window pair averaging; Biomedical imaging; Councils; Data visualization; Detectors; Electromagnetic scattering; Mirrors; Particle scattering; Positron emission tomography; Spectral shape; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference, 1992., Conference Record of the 1992 IEEE
  • Conference_Location
    Orlando, FL
  • Print_ISBN
    0-7803-0884-0
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1992.301087
  • Filename
    301087