Energy response and filter compensation of PIN Si photodiode for personal dosimetry application
Author :
Chen, Chien-Rong ; Jiang, Shiang-Huei ; Tseng, Hsun-Hua ; Chang, Hui-Fu
Author_Institution :
Dept. of Nucl. Eng., Nat. Tsing Hua Univ., Hsinchu, Taiwan
fYear :
1992
fDate :
25-31 Oct 1992
Firstpage :
70
Abstract :
The electron-photon cascade Monte Carlo code EGS4 has been used to study the energy response of the detector-grade low-cost PIN Si photodiode, which is used as the detector for personal dosimetry applications. The energy deposition distribution in the active volume of the photodiode was calculated. The effect of the discrimination level and the dimension of the active volume on the sensitivity of the photodiode was studied. The sensitivities of the photodiode with a Cu, Sn, Cu+Sn, and Pb compensation filter, respectively, were investigated in detail. The energy response of the Si photodiode was also measured in the X-ray calibration field with effective energies from 46 to 202 keV and by using Cs-137 and Co-60 γ-ray sources. Good agreement has been obtained between results of the measurement and the Monte Carlo calculation
Keywords :
Monte Carlo methods; X-ray detection and measurement; calibration; dosimeters; p-i-n photodiodes; 46 to 202 keV; Cu compensation filter; Cu-Sn compensation filter; EGS4; PIN Si photodiode; Pb compensation filter; Si; X-ray; energy response; personal dosimetry; Calibration; Detectors; Dosimetry; Filters; Leak detection; Monte Carlo methods; Noise level; Photodiodes; Silicon; Tin;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1992., Conference Record of the 1992 IEEE
Conference_Location :
Orlando, FL
Print_ISBN :
0-7803-0884-0
Type :
conf
DOI :
10.1109/NSSMIC.1992.301153
Filename :
301153
Link To Document :
بازگشت