DocumentCode :
1926468
Title :
Measurement of light output of NE213 and NE102A detectors for 2.7-14.5 MeV neutrons
Author :
Naqvi, A.A. ; Khiari, F.Z. ; Coban, A. ; Aksoy, A. ; Al-Jalal, M. Abdulaziz ; Raashid, M.
Author_Institution :
King Fahd Univ. of Pet. & Miner., Dhahran, Saudi Arabia
fYear :
1992
fDate :
25-31 Oct 1992
Firstpage :
1
Abstract :
The light output of 125-mm-diameter NE213 and NE102A detectors has been measured for neutron energies ranging from 2.7 to 14.5 MeV. For neutron energies below 6.14 MeV, measurements were carried out using the neutron time-of-flight spectrum from an Am-Be neutron source, while for proton energies above 6.14 MeV, measurements were carried out using neutrons produced from the T(d,n) reaction. For the NE102A detector the measured light output is in good agreement with the data of R.A. Cecil et al., (1979) but for the NE213 detector the light output is 2-15% lower than that for a similar detector. The NE213 detector light output agrees with the data of V. Verbinski et al. (1968)
Keywords :
neutron detection and measurement; scintillation counters; 2.7 to 14.5 MeV; NE102A detectors; NE213 detector; light output; Detectors; Electrons; Energy measurement; Gamma rays; Laboratories; Minerals; Neutrons; Petroleum; Protons; Pulse measurements;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1992., Conference Record of the 1992 IEEE
Conference_Location :
Orlando, FL
Print_ISBN :
0-7803-0884-0
Type :
conf
DOI :
10.1109/NSSMIC.1992.301165
Filename :
301165
Link To Document :
بازگشت