Title :
Compton continuum suppression by pulse shape analysis
Author :
Aspacher, B. ; Bamford, G.J. ; Coldwell, R.L. ; Rester, A.C.
Author_Institution :
Inst. for Astrophys. & Planetary Exploration, Florida Univ., Alachua, FL, USA
Abstract :
The authors present experimental results on suppressing and flattening the Compton continuum by digital and analog pulse shape analysis. A doubling of the Cs-peak-to-Co-Compton ratio and a nice flattening of the Compton edges were achieved by pulse shape analysis in rejecting pulses from the peripheral regions of the Ge-detector, where the Compton scattering followed by the gamma escaping from the detector is most common. The pulse shape analysis technique may also offer a way to construct a position-sensitive Ge-detector
Keywords :
gamma-ray detection and measurement; Compton continuum suppression; Cs-peak-to-Co-Compton ratio; Ge detector; analog; digital; position-sensitive Ge-detector; pulse shape analysis; Anodes; Astrophysics; Cathodes; Charge carrier processes; Clouds; Detectors; Extraterrestrial measurements; Pulse measurements; Pulse shaping methods; Shape measurement;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1992., Conference Record of the 1992 IEEE
Conference_Location :
Orlando, FL
Print_ISBN :
0-7803-0884-0
DOI :
10.1109/NSSMIC.1992.301182