Abstract
The authors present experimental results on suppressing and flattening the Compton continuum by digital and analog pulse shape analysis. A doubling of the Cs-peak-to-Co-Compton ratio and a nice flattening of the Compton edges were achieved by pulse shape analysis in rejecting pulses from the peripheral regions of the Ge-detector, where the Compton scattering followed by the gamma escaping from the detector is most common. The pulse shape analysis technique may also offer a way to construct a position-sensitive Ge-detector
Keywords
gamma-ray detection and measurement; Compton continuum suppression; Cs-peak-to-Co-Compton ratio; Ge detector; analog; digital; position-sensitive Ge-detector; pulse shape analysis; Anodes; Astrophysics; Cathodes; Charge carrier processes; Clouds; Detectors; Extraterrestrial measurements; Pulse measurements; Pulse shaping methods; Shape measurement;