    High sensitivity radiation detector for capillary electrophoresis

  • Author

    Gordon, J.S. ; Vasile, S. ; Hazlett, T. ; Squillante, M.

  • Author_Institution
    Radiat. Monitoring Devices, Inc., Watertown, MA, USA
  • fYear
    1992
  • fDate
    25-31 Oct 1992
  • Firstpage
    52
  • Abstract
    The authors have designed and fabricated a highly collimated, compact CdTe detector, specifically designed for measuring small radioactive sample volumes used in a capillary tube. The prototype detector was configured in a U-shape. The compact design has the advantage that it can be easily retrofitted to standard high-performance CE (capillary electrophoresis) apparatus, thereby providing additional high-sensitivity capabilities to these devices. The counting efficiency of the prototype detector was measured to be approximately 40%. Based on the detector efficiency and noise characteristics, it is estimated that the (noise-limited) minimum sample of 32P-labeled DNA fragments that can be detected is of the order of 2 attomoles. This estimate was based on the measured noise count rate (3 cpm) and the assumption that an S/N ratio of three was needed to positively identify a peak above the background
  • Keywords
    DNA; electrophoresis; phosphorus; radioactive chemical analysis; radioisotopes; semiconductor counters; 32P-labeled DNA fragments; CdTe detector; capillary electrophoresis; capillary tube; counting efficiency; radioactive sample volumes; Cadmium compounds; DNA; Electrokinetics; Fluorescence; In vivo; Instruments; Labeling; Molecular biophysics; Radiation detectors; Solid state circuits;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference, 1992., Conference Record of the 1992 IEEE
  • Conference_Location
    Orlando, FL
  • Print_ISBN
    0-7803-0884-0
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1992.301185
  • Filename
    301185