DocumentCode :
1927236
Title :
A remote gain control for photomultiplier tubes
Author :
Williams, B.E. ; Binkley, D.M. ; Casey, M.E. ; Rochelle, J.M.
Author_Institution :
CTI PET Syst., Inc., Knoxville, TN, USA
fYear :
1992
fDate :
25-31 Oct 1992
Firstpage :
245
Abstract :
Measurements were made on Burle C83062 and Hamamatsu R2497 type photomultiplier tubes (PMTs) to determine an optimum method for remotely controlling their gain. The tubes´ gain was varied by three different methods and energy and time measurements were done to determine the tube performance of each method. One method was chosen as optimum for its 10:1 gain range with only 250-ps time shift. A circuit was implemented as a computer-controllable PMT gain setting circuit which emulated this method. The circuit has applications in positron emission tomography, high energy physics, and virtually any radiation detection system involving PMTs
Keywords :
gain control; photomultipliers; telecontrol; Burle C83062; Hamamatsu R2497; PET; gain setting circuit; photomultiplier tubes; positron emission tomography; remote gain control; Cathodes; Energy measurement; Gain control; Optical coupling; Photomultipliers; Positron emission tomography; Radiation detector circuits; Resistors; Technological innovation; Voltage;