  • Abstract
    Results from the application of a modified dual photopeak window (DPW) scatter correction method to Monte Carlo simulated Tl-201 emission images are presented. In the Monte Carlo investigation, individual simulations were performed for six radiation emissions of Tl-201. For each emission, point sources of Tl-201 were imaged at various locations in a water-filled elliptical tub phantom using three energy windows: two 12% windows abutted at 72 keV and a third 10 keV window placed to the right of the photopeak window. The third window was used to estimate the spilldown contribution from the Tl-201 gamma rays in each of the two photopeak windows. Using the corrected counts in these two windows, the DPW method was applied to each point source image to estimate the scatter distribution. Application of this modified version of DPW resulted in an approximately six-fold reduction in the scatter fraction and an excellent agreement of the shape of the tails between the estimated scatter distribution and the truth
  • Keywords
    Monte Carlo methods; gamma-ray scattering; medical image processing; radioisotope scanning and imaging; 10 keV; 72 keV; 201Tl; 201Tl gamma rays; Monte Carlo simulated 201Tl emission images; dual photopeak window method; energy window; medical diagnostic imaging; nuclear medicine; scatter correction method; water-filled elliptical tub phantom; Biomedical imaging; Cancer; Gamma rays; Lakes; Medical diagnostic imaging; Medical simulation; Monte Carlo methods; Nuclear medicine; X-ray imaging; X-ray scattering;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference, 1992., Conference Record of the 1992 IEEE
  • Conference_Location
    Orlando, FL
  • Print_ISBN
    0-7803-0884-0
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1992.301481
  • Filename
    301481