    Accurate intensity correction for endorectal surface coil MR imaging of the prostate

  • Author

    Singh, Manbir ; NessAiver, Moriel

  • Author_Institution
    Dept. of Radiol. & Biomed. Eng., Univ. of Southern California, Los Angeles, CA, USA
  • fYear
    1992
  • fDate
    25-31 Oct 1992
  • Firstpage
    1307
  • Abstract
    An accurate technique has been developed to correct the intensity variation in MR (magnetic resonance) images acquired with an endorectal surface coil for prostate imaging. The technique relies on reconstructing the in-vivo size, shape, and location of the coil during imaging, wherefrom a pixel-by-pixel correction is applied to the coil´s sensitivity profile in accordance with the Biot-Savart law. The method has been evaluated by a test-object containing multiple layers of beads embedded in gelatin. Qualitatively, the improvement in visualizing regions of interest after corrections is obvious. Quantitatively, the local signal-to-noise ratio shows an improvement by a factor of 1.5-2. Several human studies have also been performed, and the technique shows remarkable improvement in visualizing the prostate region
  • Keywords
    biomedical NMR; medical image processing; Biot-Savart law; accurate intensity correction; coil sensitivity profile; endorectal surface coil MR imaging; gelatin-embedded beads; local signal-to-noise ratio; magnetic resonance imaging; medical diagnostic imaging; pixel-by-pixel correction; prostate region visualization; Coils; Image reconstruction; Magnetic resonance; Magnetic resonance imaging; Pixel; Shape; Signal to noise ratio; Surface reconstruction; Testing; Visualization;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference, 1992., Conference Record of the 1992 IEEE
  • Conference_Location
    Orlando, FL
  • Print_ISBN
    0-7803-0884-0
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1992.301519
  • Filename
    301519