• DocumentCode
    1939350
  • Title

    Development of Precise-temperature-controlled Cooling Apparatus for Medical Application by Using Peltier Effect

  • Author

    Maruyama, Shigenao ; Komiya, Atsuki ; Takeda, Hiroki ; Aiba, Setsuya

  • Author_Institution
    Inst. of Fluid Sci., Tohoku Univ., Tokyo
  • Volume
    2
  • fYear
    2008
  • fDate
    27-30 May 2008
  • Firstpage
    610
  • Lastpage
    614
  • Abstract
    This paper describes precise temperature controlled cooling apparatuses for medical application. A Peltier cryoprobe for dermis treatment and flexible cryoprobe for cryosurgery for deep region of human body were developed by utilizing the Peltier effect. The evaluation experiment of the Peltier cryoprobe was conducted. The proposed apparatus has enough performance to necrotize target diseased cells precisely. The proposed flexible cryoprobe is composed of a flexible resin tube enough to treat the deeper parts of the human body with a catheter. A Peltier element is attached at the top of the probe for active heat-transfer control of the cooling tip. The cooling experiments are carried out and the performance of the cryoprobe is evaluated. The achieved minimum temperature at the tip is -43degC.
  • Keywords
    Peltier effect; biomedical engineering; biothermics; cooling; patient treatment; probes; Peltier cryoprobe; Peltier effect; cryosurgery; dermis treatment; flexible resin tube; precise temperature controlled cooling apparatus; temperature 255 K; Biomedical equipment; Catheters; Cooling; Dermis; Humans; Medical services; Probes; Resins; Temperature control; Thermoelectricity;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    BioMedical Engineering and Informatics, 2008. BMEI 2008. International Conference on
  • Conference_Location
    Sanya
  • Print_ISBN
    978-0-7695-3118-2
  • Type

    conf

  • DOI
    10.1109/BMEI.2008.239
  • Filename
    4549247