Title :
Transmission and Reflection Diffraction Tomography in Breast Imaging
Author :
Simonetti, Francesco ; Huang, Lianjie ; Duric, Neb
Author_Institution :
Dept. of Mech. Eng., Imperial Coll. London, London
Abstract :
Commercial medical ultrasound scanners produce images of human tissue by interpreting the information carried by echoes reflected from structures contained inside it. In this paper, a new generation of toroidal ultrasound arrays is used to measure both the signals reflected and transmitted through the tissue. It is shown that transmission measurements encode complete information about the gross structure of the tissue and lead to images that are superior to those obtained form reflection measurements alone. Experimental results are provided for a gel phantom and a human breast in vivo.
Keywords :
acoustic tomography; bioacoustics; biomedical transducers; biomedical ultrasonics; mammography; phantoms; ultrasonic reflection; ultrasonic transducer arrays; ultrasonic transmission; breast imaging; gel phantom; human breast; medical ultrasound scanners; reflection diffraction tomography; reflection measurements; toroidal ultrasound arrays; transmission diffraction tomography; transmission measurements; Biomedical imaging; Breast; Diffraction; Humans; Imaging phantoms; Reflection; Signal generators; Tomography; Ultrasonic imaging; Ultrasonic variables measurement; Breast Imaging; Tomography; arrays;
Conference_Titel :
BioMedical Engineering and Informatics, 2008. BMEI 2008. International Conference on
Conference_Location :
Sanya
Print_ISBN :
978-0-7695-3118-2
DOI :
10.1109/BMEI.2008.114