    Contrast-based ultrasonic blood flow measurements based on inflow/outflow time intensities

  • Author

    Yeh, Chih-Kuang ; Wang, Sheng-Wuei ; Li, Pai-Chi

  • Author_Institution
    Dept. of Electr. Eng., Nat. Taiwan Univ., Taipei, Taiwan
  • Volume
    3
  • fYear
    2001
  • fDate
    2001
  • Firstpage
    2617
  • Abstract
    Ultrasonic contrast specific blood flow measurement techniques offer new opportunities to assess blood flow information based on evaluation of time-intensity curves (TICs). Such curves are measured to estimate the concentration of microbubbles in the blood pool. Based on the indicator-dilution theory, hemodynamic parameters such as the volume flow rate can be directly related to the time constant estimated from the TICs. In this paper, the applicability of the indicator-dilution theory is studied with an in vitro experiment setup. Moreover, the linear system assumption of the blood mixing mechanism is also tested. Several flow phantoms are constructed and a self-made, albumin based contrast agent is used. The TIC is measured by using B-mode images obtained from a commercial ultrasound system. It is found that with a bolus injection and a single mixing chamber, the estimated time constants agree with the theory despite the fact that the effective mixing volume may be smaller than the actual mixing chamber volume in some conditions. More importantly, discrepancy also exists with a prolonged injection and/or two mixing chambers with cascade connection. In other words, the linear system assumption is still questionable even under the controlled in vitro experimental conditions. Potential sources of the discrepancy require further investigation in order to develop contrast specific quantitative blood flow measurement techniques.
  • Keywords
    biomedical ultrasonics; blood flow measurement; bubbles; phantoms; B-mode images; actual mixing chamber volume; blood mixing mechanism; blood pool; bolus injection; cascade connection; commercial ultrasound system; contrast-based ultrasonic blood flow measurements; controlled in vitro experimental conditions; effective mixing volume; flow phantoms; hemodynamic parameters; in vitro experiment setup; indicator-dilution theory; inflow/outflow time intensities; linear system; microbubble concentration; prolonged injection; self-made albumin based contrast agent; single mixing chamber; time constant; time-intensity curves; volume flow rate; Blood flow; Fluid flow measurement; Hemodynamics; Imaging phantoms; In vitro; Linear systems; Measurement techniques; System testing; Time measurement; Ultrasonic variables measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2001. Proceedings of the 23rd Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-7211-5
  • Type

    conf

  • DOI
    10.1109/IEMBS.2001.1017318
  • Filename
    1017318