    Sources of error in measuring tissue pH with microsensors

  • Author

    Auerbach, Alan H. ; Soller, Babs R. ; Peura, Robert A.

  • Author_Institution
    Div. of Cardiothoracic Surg., Massachusetts Univ. Med. Center, Worcester, MA, USA
  • fYear
    1994
  • fDate
    17-18 Mar 1994
  • Firstpage
    108
  • Lastpage
    109
  • Abstract
    The use of pH microsensors in tissue is complicated by a rise in pH over a given period of time. One explanation for this occurrence is that the formation of a layer of protein on the surface of a glass electrode may affect the recorded pH. Experiments were designed to study the effect of both a uniform and nonuniform coating of fibrinogen on pH. A uniform coating is seen to affect the recorded value to a greater degree than a nonuniform coating. If a sensor does acquire a uniform layer of protein upon placement in tissue, then this error needs to be further addressed to obtain accurate pH measurements
  • Keywords
    biomedical measurement; chemical sensors; measurement errors; pH measurement; proteins; clinical measurements; error sources; glass electrode surface; microsensors; nonuniform fibrinogen coating; protein layer formation; tissue pH measurement; uniform fibrinogen coating; Biomedical measurements; Cardiology; Coatings; Electrodes; Heart; In vitro; In vivo; Microsensors; Myocardium; Proteins;