• DocumentCode
    1950970
  • Title

    Combined use of iteration, quadratic interpolation and an extra kernel for high-resolution 2D particle tracking: A first evaluation

  • Author

    Albinsson, John ; Brorsson, Sofia ; Lindgren, Finn ; Ahlgren, Åsa Rydén ; Cinthio, Magnus

  • Author_Institution
    Electr. Meas., Lund Univ., Lund, Sweden
  • fYear
    2010
  • fDate
    11-14 Oct. 2010
  • Firstpage
    2000
  • Lastpage
    2003
  • Abstract
    A novel 2D particle tracking method, that uses 1) iteration, 2) fast quadratic sub-pixel estimation (with only 28 multiplications per movement), and 3) a previous kernel, has been evaluated and compared with a full-search block-matching method. The comparison with high-frequency ultrasound data (40 MHz) was conducted in silico and on phantoms, which comprised lateral, diagonal, and ellipsoidal movement patterns with speeds of 0-15 mm/s. The mean tracking error was reduced by 68% in silico and 71% for the phantom measurements. When only sub-pixel estimation was used, the decrease in the tracking error was 61% in silico and 57% for the phantom measurements. As well as decreasing the tracking error, the new method only used 70% of the computational time needed by the full-search block-matching method. With a fast method having good tracking ability for high-frequency ultrasound data, we now have a tool to better investigate tissue movements and its dynamic functionality.
  • Keywords
    biological tissues; biomechanics; biomedical ultrasonics; data analysis; interpolation; iterative methods; medical computing; phantoms; biomedical imaging; diagonal movement patterns; ellipsoidal movement patterns; fast quadratic subpixel estimation; frequency 40 MHz; full-search block-matching method; high-frequency ultrasound data; high-resolution 2D particle tracking; in-silico analysis; iterative method; lateral movement patterns; mean tracking error; phantoms; quadratic interpolation; tissue movements; Estimation; Kernel; Phantoms; Pixel; Tracking; Ultrasonic imaging; Ultrasonic variables measurement; Adaptive rood pattern search; Block-matching; Tissue motion;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2010 IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1948-5719
  • Print_ISBN
    978-1-4577-0382-9
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2010.5935430
  • Filename
    5935430