Title :
A preliminary study on photoplethysmogram (PPG) signal analysis for reduction of motion artifact in frequency domain
Author :
Cho, J.M. ; Sung, Y.K. ; Shin, K.W. ; Jung, D.J. ; Kim, Yong Sin ; Kim, Nicholas H.
Author_Institution :
Dept. of Biomed. Eng., Inje Univ., Gimhae, South Korea
Abstract :
In this study, we analyzed frequency characteristics of photoplethysmogram (PPG) signal to improve the accuracy of SpO2 by reducing the motion artifacts. 23 kinds of different signals generated by SpO2 test equipment were used for this analysis. The results of analysis showed that most of PPG signals analyzed in this study consisted of the fundamental, 2nd and 3rd harmonic components caused by arterial blood flow and component caused by motion artifact. The result also showed the relative magnitudes of its three frequency components. The signals could be divided into four groups based on the relative locations of these frequency components. But four different algorithms would be required to discriminate the signal and motion artifact frequency components.
Keywords :
biomedical equipment; blood; blood vessels; haemodynamics; medical signal detection; medical signal processing; photoplethysmography; PPG signal analysis; SpO2 test equipment; arterial blood flow; frequency characteristics; frequency domain; motion artifact reduction; photoplethysmogram signal analysis; second harmonic components; third harmonic components; Motion artifact; PPG; SpO2;
Conference_Titel :
Biomedical Engineering and Sciences (IECBES), 2012 IEEE EMBS Conference on
Conference_Location :
Langkawi
Print_ISBN :
978-1-4673-1664-4
DOI :
10.1109/IECBES.2012.6498141