• DocumentCode
    1962628
  • Title

    On Automatic Measurement of Microwave Noise Figure

  • Author

    Grabowski, Krzysztof ; Adamski, Wladyslaw ; Dabrowski, Jerzy

  • Author_Institution
    Technical University of Gda?sk, Telecommunication Institute, 80-952 GDANSK, POLAND
  • fYear
    1974
  • fDate
    10-13 Sept. 1974
  • Firstpage
    76
  • Lastpage
    80
  • Abstract
    In conventional microwave automatic measurements of the amplifier noise figure the measurement results are distorted by an error which is due to the noise of the receiver itself which is connected in cascade with the amplifier under test. This error depends on the receiver noise and on the gain of the amplifier. In the paper there is presented a method which eliminates this error and a set-up is described which proves in practice the advantages of the method worked out. In the method - besides of the two measurement states in which the noise generator is "on" and "off" - there is included a third state in which the receiver own noise level is measured. This is enable due to an electronically controlled microwave p-i-n diode switch which -on the other side- does not affect the overall set-up accuracy.
  • Keywords
    Automatic testing; Distortion measurement; Microwave amplifiers; Microwave measurements; Noise figure; Noise generators; Noise level; Noise measurement; P-i-n diodes; Switches;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microwave Conference, 1974. 4th European
  • Conference_Location
    Montreux, Switzerland
  • Type

    conf

  • DOI
    10.1109/EUMA.1974.332015
  • Filename
    4130656