Title :
Numerical design and evaluation of Vector Doppler Ultrasound systems for accurate 3-D velocity measurements
Author :
Hussain, B. ; Wong, E.Y. ; Poepping, T.L.
Author_Institution :
Depts. of Phys. & Astron., Univ. of Western Ontario, London, ON, Canada
Abstract :
Vector Doppler ultrasound (VDUS) systems offer the potential for improved accuracy in mapping of complex flow parameters, such as recirculation, turbulence and shear stress, which are probable risk factors leading to vascular disease and stroke. VDUS system configurations with 2 to 6 receivers and corresponding reconstructions of a complex velocity field were numerically evaluated.
Keywords :
Doppler effect; biomedical ultrasonics; blood flow measurement; diseases; velocity measurement; 3D velocity measurement; VDUS system configuration; flow parameters; risk factors; stroke; vascular disease; vector Doppler ultrasound system; Acoustics; Carotid arteries; Computational fluid dynamics; Doppler effect; Receivers; Ultrasonic imaging; Velocity measurement; blood velocity; carotid artery; hemodynamics; vector Doppler ultrasound;
Conference_Titel :
Ultrasonics Symposium (IUS), 2010 IEEE
Conference_Location :
San Diego, CA
Print_ISBN :
978-1-4577-0382-9
DOI :
10.1109/ULTSYM.2010.5936012