This paper outlines the evolution of the stand alone cavity auto tuning of the active hydrogen maser using microwave cavity frequency switching technique. Analysis of basic destabilizing factors limiting short- term, long- term frequency stability and phase noise of output signal of this technique is given. The effect of cavity tuning on short and medium term stability characteristics is negligible. Long-term frequency stability of (3÷5) × 10-16 at averaging time 1÷20 days has been achieved.