DocumentCode
1965127
Title
A New Procedure for System Calibration and Error Removal in Automated S-Parameter Measurements
Author
Franzen, Norman R. ; Speciale, Ross A.
fYear
1975
fDate
1-4 Sept. 1975
Firstpage
69
Lastpage
73
Abstract
A new procedure is described, intended for system calibration and systematic error removal in automated S-parameter measurements. The new procedure uses only three non-critical reference standards, called the "Through", the "Short" and the "Delay" and leads to a set of equations for which a fast, non-iterative, explicit solution is known.
Keywords
Automatic testing; Calibration; Delay; Equations; Frequency; Measurement standards; Performance evaluation; Scattering parameters; Standards development; System testing;
fLanguage
English
Publisher
ieee
Conference_Titel
Microwave Conference, 1975. 5th European
Conference_Location
Hamburg, Germany
Type
conf
DOI
10.1109/EUMA.1975.332155
Filename