• DocumentCode
    1977991
  • Title

    A method to determine diastolic blood pressure based on pressure pulse propagation in the electronic palpation method

  • Author

    Sorvoja, H.S.S. ; Myllylä, R.A. ; Nissilä, S. ; Kärjä-Koskenkari, P.M. ; Koskenkari, J.K. ; Lilja, M.K. ; Kesaniemi, Y.A.

  • Author_Institution
    Dept. of Electr. Eng., Oulu Univ., Finland
  • Volume
    4
  • fYear
    2001
  • fDate
    2001
  • Firstpage
    3255
  • Abstract
    Evaluates the accuracy of the pulse transit time method for determining diastolic blood pressure using intra-arterial blood pressure as a reference. First, the paper describes the method. Then it discusses two sets of measurements which were carried out to determine the accuracy of the method with cardiac operated patients and healthy young volunteers. The thus obtained accuracies were +0.7 mmHg ± 10.7 mmHg for the cardiac patients and -6.6 mmHg ± 10.5 mmHg for the young volunteers, respectively.
  • Keywords
    biomedical transducers; blood pressure measurement; patient monitoring; pressure transducers; accuracy; cardiac operated patients; diastolic blood pressure; electronic palpation method; healthy young volunteers; intra-arterial blood pressure; multi-element transducer array; pressure pulse propagation; pulse transit time method; Arteries; Biomedical monitoring; Blood pressure; Brachytherapy; Fingers; Hospitals; Pressure measurement; Pulse measurements; Testing; Transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2001. Proceedings of the 23rd Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-7211-5
  • Type

    conf

  • DOI
    10.1109/IEMBS.2001.1019518
  • Filename
    1019518