Abstract
Evaluates the accuracy of the pulse transit time method for determining diastolic blood pressure using intra-arterial blood pressure as a reference. First, the paper describes the method. Then it discusses two sets of measurements which were carried out to determine the accuracy of the method with cardiac operated patients and healthy young volunteers. The thus obtained accuracies were +0.7 mmHg ± 10.7 mmHg for the cardiac patients and -6.6 mmHg ± 10.5 mmHg for the young volunteers, respectively.
Keywords
biomedical transducers; blood pressure measurement; patient monitoring; pressure transducers; accuracy; cardiac operated patients; diastolic blood pressure; electronic palpation method; healthy young volunteers; intra-arterial blood pressure; multi-element transducer array; pressure pulse propagation; pulse transit time method; Arteries; Biomedical monitoring; Blood pressure; Brachytherapy; Fingers; Hospitals; Pressure measurement; Pulse measurements; Testing; Transducers;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2001. Proceedings of the 23rd Annual International Conference of the IEEE
ISSN
1094-687X
Print_ISBN
0-7803-7211-5
Type
conf
DOI
10.1109/IEMBS.2001.1019518
Filename
1019518
Link To Document