• DocumentCode
    1978161
  • Title

    Active shielding in measurements of DC near biomagnetic fields

  • Author

    Nowak, H. ; Haueisen, J. ; Ziolkowski, M. ; Resmer, F. ; Schüler, J. ; Giessler, F.

  • Volume
    4
  • fYear
    2001
  • fDate
    2001
  • Firstpage
    3277
  • Abstract
    Measurements of DC near biomagnetic fields are disturbed by low frequency noise, that is not reduced sufficiently by most magnetically shielded rooms or gradiometers. An effective SQUID based active shielding system has been developed and installed at the magnetically shielded rooms in the Biomagnetic Center of the University Jena to reduce external low frequency disturbances. A reduction of the magnetic noise of about 23 dB could be achieved at 50 mHz.
  • Keywords
    SQUID magnetometers; biomagnetism; magnetic field measurement; magnetic shielding; magnetoencephalography; 50 mHz; DC near biomagnetic field measurements; active shielding; effective SQUID based active shielding system; external low frequency disturbances; gradiometers; human stroke patients; low frequency noise; magnetic noise; magnetic source imaging; magnetically shielded rooms; periinfarct depolarizations; spreading depression; Biomagnetics; Frequency; Low-frequency noise; Magnetic field measurement; Magnetic noise; Magnetic recording; Magnetic shielding; Noninvasive treatment; SQUIDs; Superconducting magnets;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2001. Proceedings of the 23rd Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-7211-5
  • Type

    conf

  • DOI
    10.1109/IEMBS.2001.1019524
  • Filename
    1019524