DocumentCode :
1978203
Title :
The Fringing Capacitance of Reference Open Circuits Used in Network Analyser Calibration
Author :
Ambrose, C. ; Aitchison, C.S.
Author_Institution :
Department of Electronics, Chelsea College, University of London, Pulton Place, London SW6 5PR, ENGLAND
fYear :
1980
fDate :
8-12 Sept. 1980
Firstpage :
273
Lastpage :
277
Abstract :
The effect of incorrectly calibrating an automatic network analyser by using a wrong value for the open-circuit fringing capacitance has been studied. It was found that any measurement made using this calibration will be in error by an amount that is independent of the network analyser error network. A new linear frequency dependent value for the fringing capacitance of an APC-7 shielded open-circuit has been established.
Keywords :
Calibration; Capacitance measurement; Circuits; Differential equations; Error analysis; Error correction; Frequency measurement; Impedance measurement; Intelligent networks; Reflection;