• DocumentCode
    1983059
  • Title

    Automated method for Doppler echocardiography image analysis

  • Author

    Shechner, Oron ; Scheinowitz, M. ; Feinberg, Micha S. ; Greenspan, H.

  • Author_Institution
    Dept. of Biomed. Eng., Tel Aviv Univ., Israel
  • fYear
    2004
  • fDate
    6-7 Sept. 2004
  • Firstpage
    177
  • Lastpage
    180
  • Abstract
    An automated method using image processing and computer modeling tools has been developed for the extraction of clinical data from Doppler echocardiography images. Currently, medical data from Doppler images is extracted manually and is subject to large inter- and intra-observer variability. An automated Doppler analysis method is important for improving the analysis accuracy and robustness, especially for patients with abnormal heart-rates. The presented methodology works on signals from the mitral valve (MV) and the tricuspid valve (TV), and detects the envelope of the velocity versus time function, the maximal velocity envelope (MVE). From the MVE, clinical parameters are extracted (peak velocity, area beneath the curve). Experimental results show a strong correlation between the automatically extracted parameters and parameters acquired manually.
  • Keywords
    Doppler measurement; echocardiography; medical image processing; Doppler echocardiography images; abnormal heart-rates; automated method; computer modeling tools; envelope detection; image analysis; image processing; maximal velocity envelope; mitral valve; patients; tricuspid valve; velocity versus time function; Algorithm design and analysis; Biomedical imaging; Data mining; Echocardiography; Electrocardiography; Heart; Image analysis; Image processing; TV; Valves;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electrical and Electronics Engineers in Israel, 2004. Proceedings. 2004 23rd IEEE Convention of
  • Print_ISBN
    0-7803-8427-X
  • Type

    conf

  • DOI
    10.1109/EEEI.2004.1361118
  • Filename
    1361118