DocumentCode
1984571
Title
The Choice of Sliding Load Positions to Improve Network Analyser Calibration
Author
Heyker, H.C.
Author_Institution
Twente University of Technology, EL/Microwave Group, P.O. Box 217, 7500 AE ENSCHEDE. Netherlands.
fYear
1982
fDate
13-17 Sept. 1982
Firstpage
429
Lastpage
434
Abstract
Using a sliding load for the calibration of a network analyser a wrong choice of sliding load positions can cause an incorrect calibration and lead to errors in the corrected measurements of an unknown reflection coefficient. The frequency dependence and error sensitivity have been studied for two distributions of the sliding load positions which are both easy to use. Application formulas are presented for the sliding load used with these data point distributions.
Keywords
Calibration; Covariance matrix; Error analysis; Error correction; Frequency dependence; Frequency measurement; Least squares approximation; Position measurement; Reflection; Scattering parameters;