DocumentCode :
1989296
Title :
Characterization and Measurement of Radiometer Stability
Author :
Rau, G. ; Schieder, R. ; Vowinkel, B.
Author_Institution :
Physikalisches Institut der Universitÿt Köln, UniversitÿtsstraÃ\x9fe 14
fYear :
1984
fDate :
10-13 Sept. 1984
Firstpage :
248
Lastpage :
253
Abstract :
For the characterization of radiometer noise and stability a new and very practical method is introduced. It consists of the measurement of the "Allan-variance" well known for the characterization of the stability of frequency standards. The plot of the Allan variance versus integration time allows one to determine the different types of fluctuations of the radio-meter output signal. In particular, the best range of integration time for optimum use of the system can accurately be evaluated. Theoretical considerations and experimental results are presented.