• DocumentCode
    1996367
  • Title

    Design of a blood pump for a wearable artificial kidney device

  • Author

    Markovic, Miroslav ; Rapin, Michael ; Correvon, Marc ; Perriard, Yves

  • Author_Institution
    Integrated Actuators Lab. (LAI), EPFL, Lausanne, Switzerland
  • fYear
    2012
  • fDate
    15-20 Sept. 2012
  • Firstpage
    3649
  • Lastpage
    3654
  • Abstract
    The aim of a European project Nephron+ is design of a wearable artificial kidney device (WAKD). This paper is focused on design of the corresponding ultralow-hemolysis continuous-operation blood pump. Accurate specifications and operating principle of the pump are determined; a first, non-optimal configuration of an electro-mechanical actuator which will be used to pump the blood is designed; its prototype is presented along with the corresponding driving electronic circuit; finally, based on the measurements, the actuator is optimized, and the final design and first results are presented.
  • Keywords
    artificial organs; biomedical electronics; blood; electromechanical actuators; haemodynamics; kidney; pumps; European project; Nephron+; WAKD; blood pump design; electro-mechanical actuator; electronic circuit; non-optimal configuration; ultralow-hemolysis continuous-operation blood pump; wearable artificial kidney device; Actuators; Analytical models; Blood; Electron tubes; Force; Prototypes; Temperature measurement; Actuators; Efficiency; Hemolysis; Medical control systems; Microcontroller; Pumps;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Energy Conversion Congress and Exposition (ECCE), 2012 IEEE
  • Conference_Location
    Raleigh, NC
  • Print_ISBN
    978-1-4673-0802-1
  • Electronic_ISBN
    978-1-4673-0801-4
  • Type

    conf

  • DOI
    10.1109/ECCE.2012.6342308
  • Filename
    6342308