Abstract
In Dual Head Coincidence Imaging (DHCI), rotational weights must be applied to detected coincidence events to compensate for varying dwell-times. The dwell-time depends on both the transaxial and axial distance coordinates describing a coincidence line in parallel projection space. An exact formula is derived for the dwell-time of a DHCI camera with continuous detector rotation given a coincidence line described in parallel projection coordinates. The formula is verified by simulation. In addition, the boundary of the region where no data are measured is described. This missing-data region differs from that of a cylindrical PVI system whose diameter is the same as the head separation of the DHCI system and whose axial extent is the same as the axial dimension of the rotating planar detectors in the DHCI system
Keywords
biomedical equipment; cameras; positron emission tomography; axial distance coordinates; coincidence line; exact formula; medical diagnostic imaging; medical instrumentation; missing-data region; nuclear medicine; parallel projection coordinates; parallel projection space; rotating planar detectors; rotational weights; transaxial distance coordinates; varying dwell-times compensation; Azimuthal angle; Cameras; Collimators; Detectors; Event detection; Geometry; Magnetic heads; Optical imaging; Shape;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
Conference_Location
Seattle, WA
ISSN
1082-3654
Print_ISBN
0-7803-5696-9
Type
conf
DOI
10.1109/NSSMIC.1999.842772
Filename
842772
Link To Document