Title
Optimization of input parameters for a segmented attenuation algorithm for the GE advanceTM scanner
Author
Ross, S. ; Crandall, P. ; Stearns, C. ; Bullock, M. ; Bettinardi, V. ; Pagani, E. ; Lewellen, T. ; Kohlmyer, S.
Author_Institution
GE Med. Syst., Waukesha, WI, USA
Volume
3
fYear
1999
fDate
1999
Firstpage
1632
Abstract
Segmented attenuation correction (SAG) techniques provide the possibility for increasing patient throughput by decreasing transmission scan times from those required for standard measured attenuation correction (MAC) methods. A SAC algorithm using an unsupervised least biased fuzzy clustering method has been implemented on the GE Advance TM system. This implementation has the advantage of requiring no a priori region definitions. There are, however, several parameters which affect both the bias and variance produced by the SAC algorithm. The authors used statistical design of experiment (DOE) techniques to determine the subset of critical input parameters that have the largest impact on both bias and variance. These critical parameters were then analyzed to determine their influence on bias and variance. A second DOE was then carried to optimize the remaining parameters. Settings are identified which can be used to reduce both bias and variance in the segmentation algorithm
Keywords
biomedical equipment; electromagnetic wave absorption; image segmentation; medical image processing; optimisation; GE advance scanner; a priori region definitions; critical input parameters subset; input parameters optimization; medical diagnostic imaging; patient throughput increase; segmentation algorithm; segmented attenuation algorithm; statistical design of experiment techniques; Attenuation; Clustering algorithms; Filtering; Filters; Histograms; Image reconstruction; Image segmentation; Measurement standards; Throughput; US Department of Energy;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
Conference_Location
Seattle, WA
ISSN
1082-3654
Print_ISBN
0-7803-5696-9
Type
conf
DOI
10.1109/NSSMIC.1999.842886
Filename
842886
Link To Document