Title :
Double-tuned RF receiver coil for detecting both 1H and 31P elements in 4.7 T MRI system
Author :
Sheikh, Faisal Ahmad ; Young Cheol Kim ; Ick Chang Choi ; Hyun Deok Kim
Author_Institution :
Sch. of Electron. Eng., Kyungpook Nat. Univ., Daegu, South Korea
Abstract :
A novel double-tuned radio-frequency (RF) receiver coil for use in magnetic resonance imaging (MRI) systems has been demonstrated based on a birdcage-type circuit. The receiver coil utilises the end-ring current as well as the leg current of the birdcage-type circuit to detect the nuclear magnetic resonance (NMR) signals of different elements simultaneously. It was designed to detect the NMR signals of both 1H and 31P elements in a 4.7 T MRI system. The coil showed an excellent frequency response and provided clear magnetic resonance images of a Na2HPO4 phantom when used with the 4.7 T MRI system.
Keywords :
biomedical MRI; coils; MRI system; birdcage type circuit; double tuned RF receiver coil; end ring current; magnetic flux density 4.7 T; magnetic resonance imaging; nuclear magnetic resonance signals;
Journal_Title :
Electronics Letters
DOI :
10.1049/el.2015.1086