  • Abstract
    There are many advantages to developing an algorithm which will extend calibration intervals of Automatic Test Equipment (ATE). A historical record must be kept to demonstrate that the equipment remains within specifications without additional calibration. The calibration extending algorithm must reject noise, and account for variations due to temperature. A new method, which extrapolates expected variations caused by temperature changes and anticipates the variations caused by noise, is described and analyzed. Experimental results are given
  • Keywords
    automatic test equipment; calibration; interference suppression; stability; ATE; automatic test equipment; calibration extending algorithm; noise rejection; temperature variations; Calibration; Clocks; Instruments; Stability; Switches; Temperature measurement; Temperature sensors; Time measurement; Uncertainty; Voltage;