DocumentCode
2034341
Title
Extending calibration intervals
Author
Max, Solomon
Author_Institution
LTX Corp., Westwood, MA, USA
fYear
1996
fDate
20-25 Oct 1996
Firstpage
118
Lastpage
126
Abstract
There are many advantages to developing an algorithm which will extend calibration intervals of Automatic Test Equipment (ATE). A historical record must be kept to demonstrate that the equipment remains within specifications without additional calibration. The calibration extending algorithm must reject noise, and account for variations due to temperature. A new method, which extrapolates expected variations caused by temperature changes and anticipates the variations caused by noise, is described and analyzed. Experimental results are given
Keywords
automatic test equipment; calibration; interference suppression; stability; ATE; automatic test equipment; calibration extending algorithm; noise rejection; temperature variations; Calibration; Clocks; Instruments; Stability; Switches; Temperature measurement; Temperature sensors; Time measurement; Uncertainty; Voltage;
fLanguage
English
Publisher
ieee
Conference_Titel
Test Conference, 1996. Proceedings., International
Conference_Location
Washington, DC
ISSN
1089-3539
Print_ISBN
0-7803-3541-4
Type
conf
DOI
10.1109/TEST.1996.556953