Estimation of blood pressure using Photoplethysmography on the wrist
Author
Song, S.H. ; Cho, J.S. ; Oh, H.S. ; Lee, J.S. ; Kim, I.Y.
Author_Institution
Dept. of Biomed. Eng., Hanyang Univ., Seoul, South Korea
fYear
2009
fDate
13-16 Sept. 2009
Firstpage
741
Lastpage
744
Abstract
Blood pressure measuring method using PPG signal is the one of many non-invasive blood pressure methods. In this study, we have studied a new method that estimates the blood pressure using wrist cuff and wrist PPG (photoplethysmograph) signal. During the deflation of the wrist cuff pressure, the PPG pulse appears at certain point that is similar to Korotkoff sound. After the pulse appeared, the morphology of PPG pulses was changed into the certain shape. So, we use these points to estimate the blood pressure. For validation, the reference SBP and DBP are measured by auscultatory method at the left upper arm. Subjects were 13 non cardiac disease of man in their twenties. The result of this study, the mean difference of SBP and DBP, between auscultatory method and PPG analysis is 2.08 ± 5.07 mmHg and 1.38 ± 7.86 mmHg respectively.
Keywords
blood pressure measurement; plethysmography; Korotkoff sound; auscultatory method; blood pressure estimation; photoplethysmography; wrist PPG signal; wrist cuff; Biomedical measurements; Biomedical monitoring; Blood pressure; Control systems; Morphology; Pressure control; Pressure measurement; Pulse shaping methods; Sensor arrays; Wrist;
fLanguage
English
Publisher
ieee
Conference_Titel
Computers in Cardiology, 2009
Conference_Location
Park City, UT
ISSN
0276-6547
Print_ISBN
978-1-4244-7281-9
Electronic_ISBN
0276-6547
Type
conf
Filename
5445278
Link To Document