Title :
A CMOS turbidity to frequency converter with calibration circuits for detecting turbidity applications
Author :
Chiang, Cheng-Ta ; Huang, Shih-Ming
Author_Institution :
Department of Electrical Engineering, National Chia Yi University, Taiwan
Abstract :
In this paper, a CMOS turbidity to frequency converter with calibration circuits for detecting turbidity applications is newly proposed. A calibration technique is proposed in this work. Another innovation is that the outputs of the proposed chip are directly digitized, they could be easily sent over a wide range of transmission media, such as PSN, radio, optical, IR, ultrasonic, and etc. Before performing the proposed calibration circuits, the maximum linear error is 25.82%. After calibration, the maximum linear error is reduced to 0.89%. The turbidity ranges from 1.5 to 992 Nephelometric Turbidity Units (NTU), and its corresponding output frequency range is 24.875 to 18.214 kHz. The chip area is 1.01 × 1.01 mm2. The proposed chip is suitable for devices of measuring water turbidity.
Keywords :
CMOS integrated circuits; Calibration; Capacitance; Frequency conversion; Optical sensors; Robot sensing systems; Simulation; calibration circuits; turbidity sensor; water turbidity;
Conference_Titel :
Mechatronics and Automation (ICMA), 2015 IEEE International Conference on
Conference_Location :
Beijing, China
Print_ISBN :
978-1-4799-7097-1
DOI :
10.1109/ICMA.2015.7237515