• DocumentCode
    2047837
  • Title

    Analyzing the dynamic range of fluorescent detection in microarray scanner

  • Author

    Liqian, Wang ; Zukang, Lu

  • Author_Institution
    Nat. Eng. & Technol. Res. Center for Opt. Instrum., Zhejiang Univ., Hangzhou
  • fYear
    2006
  • fDate
    16-18 Oct. 2006
  • Firstpage
    136
  • Lastpage
    138
  • Abstract
    By analyzing the dark current noise of photomultiplier tube (PMT) and the thermal noise of feedback resistor, the detection limit of the processing circuit in the microarray scanner is calculated theoretically, and the saturated signal of that is obtained by the anode radiant sensitivity of the PMT under optimized gain. The dynamic range is the saturated signal divided by the detection limit, which is 8773 in theory. The experiment result was acquired after scanning a fluorescent calibration slide by the scanner, which was 4096.
  • Keywords
    biochemistry; fluorescence; lab-on-a-chip; molecular biophysics; photomultipliers; proteins; DNA microarray; anode radiant sensitivity; dark current noise; feedback resistor; fluorescent calibration slide; fluorescent detection; microarray scanner; photomultiplier tube; protein microarray; thermal noise; Circuit noise; Dark current; Dynamic range; Feedback circuits; Fluorescence; Photomultipliers; Resistors; Signal analysis; Signal processing; Thermal resistance;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biophotonics, Nanophotonics and Metamaterials, 2006. Metamaterials 2006. International Symposium on
  • Conference_Location
    Hangzhou
  • Print_ISBN
    0-7803-9773-8
  • Electronic_ISBN
    0-7803-9774-6
  • Type

    conf

  • DOI
    10.1109/METAMAT.2006.335017
  • Filename
    4134756