DocumentCode
2047837
Title
Analyzing the dynamic range of fluorescent detection in microarray scanner
Author
Liqian, Wang ; Zukang, Lu
Author_Institution
Nat. Eng. & Technol. Res. Center for Opt. Instrum., Zhejiang Univ., Hangzhou
fYear
2006
fDate
16-18 Oct. 2006
Firstpage
136
Lastpage
138
Abstract
By analyzing the dark current noise of photomultiplier tube (PMT) and the thermal noise of feedback resistor, the detection limit of the processing circuit in the microarray scanner is calculated theoretically, and the saturated signal of that is obtained by the anode radiant sensitivity of the PMT under optimized gain. The dynamic range is the saturated signal divided by the detection limit, which is 8773 in theory. The experiment result was acquired after scanning a fluorescent calibration slide by the scanner, which was 4096.
Keywords
biochemistry; fluorescence; lab-on-a-chip; molecular biophysics; photomultipliers; proteins; DNA microarray; anode radiant sensitivity; dark current noise; feedback resistor; fluorescent calibration slide; fluorescent detection; microarray scanner; photomultiplier tube; protein microarray; thermal noise; Circuit noise; Dark current; Dynamic range; Feedback circuits; Fluorescence; Photomultipliers; Resistors; Signal analysis; Signal processing; Thermal resistance;
fLanguage
English
Publisher
ieee
Conference_Titel
Biophotonics, Nanophotonics and Metamaterials, 2006. Metamaterials 2006. International Symposium on
Conference_Location
Hangzhou
Print_ISBN