• DocumentCode
    2050856
  • Title

    VUV Si photodiode as the gas proportional scintillation counter readout

  • Author

    Lopes, J.A.M. ; Fernandes, L.M.P. ; dos Santos, J.M.F. ; Conde, C.A.N.

  • Author_Institution
    Inst. Superior de Engenharia de Coimbra, Portugal
  • Volume
    2
  • fYear
    1999
  • fDate
    24-30 Oct. 1999
  • Firstpage
    646
  • Abstract
    The application of recently developed VUV silicon photodiodes as a replacement for the standard photomultiplier tube (PMT) in a gas proportional scintillation counter (GPSC) is investigated. A photodiode positioned within the gas envelope would result in a GPSC that is more compact, robust, and simpler to operate. At the 170 nm wavelength of the scintillation light from xenon, UV photodiodes achieve an effective quantum efficiency of 72%, better than the typically 20% figure for PMTs. On the other hand, the lack of amplification in the photodiode readout limits the signal-to-noise ratio that can be achieved. A GPSC based on a silicon photodiode with an active area of 100 mm/sup 2/ and no additional amplification stage achieved energy resolutions of 20.5, 14.5 and 13.4% for 5.9, 22.1, and 59.6 keV X-rays, respectively. Signal noise in the low energy limit extends up to approximately 3 keV.
  • Keywords
    gas scintillation detectors; photodiodes; silicon; xenon; Si; VUV Si photodiode; Xe; effective quantum efficiency; energy resolution; gas proportional scintillation counter readout; low energy limit; photodiode readout; scintillation light; signal-to-noise ratio; Energy resolution; Photodiodes; Photomultipliers; Robustness; Scintillation counters; Signal to noise ratio; Silicon; Standards development; X-rays; Xenon;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
  • Conference_Location
    Seattle, WA, USA
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-5696-9
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1999.845752
  • Filename
    845752