DocumentCode
2051042
Title
The change of gamma equivalent energy with temperature for scintillation detector assemblies
Author
Rozsa, C. ; Grodsinsky, C. ; Penn, D. ; Raby, P. ; Schreiner, R.
Author_Institution
Bicron, Newbury, OH, USA
Volume
2
fYear
1999
fDate
1999
Firstpage
686
Abstract
The change in the relative light output of the alpha and gamma response of scintillators with temperature has been measured. BGO, CaF 2(Eu), NaI(Tl) and YAP(Ce) were investigated over the temperature range from -20°C to 70°C. The GEE and the normalized light output for alpha particles from Am241 and for gamma rays from Na22 are presented in tabular and graphical forms
Keywords
alpha-particle detection; bismuth compounds; calcium compounds; cerium; europium; gamma-ray detection; sodium compounds; solid scintillation detectors; thallium; yttrium compounds; Bi4Ge3O12; CaF2:Eu; NaI:Tl; YAP:Ce; YAlO3:Ce; alpha response; gamma equivalent energy; gamma response; normalized light output; relative light output; scintillation detector assemblies; Alpha particles; Assembly; Ceramics industry; Fires; Gamma rays; Rubber; Scintillation counters; Temperature control; Temperature distribution; Temperature measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
Conference_Location
Seattle, WA
ISSN
1082-3654
Print_ISBN
0-7803-5696-9
Type
conf
DOI
10.1109/NSSMIC.1999.845761
Filename
845761
Link To Document