Title :
Uranium pinhole collimators for positron emission SPECT imaging of small volumes
Author :
Tenney, C.R. ; Tornai, M.P. ; Smith, M.F.
Author_Institution :
Duke Univ., Durham, NC
Abstract :
Depleted uranium (DU) pinhole collimators are evaluated for small volume SPECT imaging of 511 keV gammas from F-18 positrons, with a view towards small animal imaging. Tungsten (W) pinholes are examined for comparison. Line source measurements show FWHM as low as 6.5 mm for a 1 mm pinhole, and SPECT system FWHM of 5.2 mm at 39 mm ROR, and 6.3 mm at 69 mm ROR. Planar FWHM is smaller for DU than for W pinholes but is insensitive to aperture size for 1-3 mm dia. pinholes. These trends indicate penetration at pinhole edges, as confirmed by LSF shape. In test scans, neither DU nor W pinholes were able to resolve cold rods as large as 4.7 mm (diameter=2 X center-to-center spacing). U pinholes resolve 3.9 mm hot rods, imaging either F-18 alone or imaging F-18 simultaneously with Tc-99m. (Tc-99m scans, either alone or with FDG, resolved 2.3 mm rods.) DU pinhole SPECT currently shows poorer spatial resolution than high-resolution PET, but is open to improvements, and is capable of multiple-isotope acquisitions
Keywords :
biomedical equipment; image reconstruction; image resolution; medical image processing; single photon emission computed tomography; uranium; 511 keV; U; cold rod phantom; depleted uranium pinhole collimators; dual-isotope scan; image reconstruction; multiple-isotope acquisitions; penetration at pinhole edges; planar line source data; positron emission SPECT imaging; small animal imaging; small volumes; spatial resolution; Animals; Apertures; Collimators; High-resolution imaging; Image resolution; Optical imaging; Positrons; Radioactive decay; Spatial resolution; Tungsten;
Conference_Titel :
Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
Conference_Location :
Seattle, WA
Print_ISBN :
0-7803-5696-9
DOI :
10.1109/NSSMIC.1999.845839