DocumentCode :
2061970
Title :
HTSC SQUID-based set-up for calibration of voltage dividers and AC voltmeters
Author :
Karpov, O.V. ; Gumenuk, V.S. ; Koutovoi, V.D. ; Shilov, K.A. ; Shakirzyanova, F.R.
Author_Institution :
Nat. Res. Inst. for Phys.-Tech. & Radiotech. Meas., Mendeleevo, Russia
Volume :
1
fYear :
1996
fDate :
1996
Firstpage :
220
Abstract :
HTSC SQUID made of HTSC ceramics has been fabricated and studied. The SQUID has been used in the set-up for calibrating AC voltmeters and voltage dividers. The measurement technique is based on strict periodical dependence of a SQUID voltage-flux characteristic on a magnetic flux. The dynamic range of the set-up was from 300 μV to 10 V. The frequency range was from 100 kHz to 1 MHz. The calibration error was about 0001 dB. The SQUID has been used in the laboratory since 1992
Keywords :
SQUID magnetometers; calibration; high-temperature superconductors; voltage dividers; voltage measurement; voltmeters; 100 kHz to 1 MHz; 300 muV to 10 V; AC voltmeters; HTSC SQUID; HTSC ceramics; calibration; calibration error; dynamic range; frequency range; laboratory; magnetic flux; periodical dependence; voltage dividers; voltage-flux characteristic; Calibration; Ceramics; Dynamic range; Frequency; Laboratories; Magnetic flux; Measurement techniques; SQUIDs; Voltage; Voltmeters;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Instrumentation and Measurement Technology Conference, 1996. IMTC-96. Conference Proceedings. Quality Measurements: The Indispensable Bridge between Theory and Reality., IEEE
Conference_Location :
Brussels
Print_ISBN :
0-7803-3312-8
Type :
conf
DOI :
10.1109/IMTC.1996.507380
Filename :
507380
Link To Document :
بازگشت