    A SAR-probe calibration system using reference dipole antenna in tissue-equivalent liquid

  • Author

    Hamada, Lira ; Sato, Kenichi ; Ishii, Nozomu ; Watanabe, Soichi

  • Author_Institution
    Nat. Inst. of Inf. & Commun. Technol., Tokyo
  • fYear
    2008
  • fDate
    19-23 May 2008
  • Firstpage
    116
  • Lastpage
    119
  • Abstract
    The SAR-probe is usually calibrated using a rectangular waveguide with a matching dielectric spacer below 3 GHz. However, the probe diameter is comparable with the cross-section area of the waveguide above 3 GHz and it can deteriorate the accuracy of the calibration. Therefore, there is an alternative method for calibrating the SAR-probe; relating the output voltage of the probe to the field intensity produced by a reference antenna in the tissue-equivalent liquid by two-antenna method. In this report, we have built a prototype of a calibration system and presented some calibration results of SAR-probe sensitivity using this method in conducting head-equivalent liquid at 2.45 GHz.
  • Keywords
    biological effects of microwaves; calibration; dipole antennas; electromagnetic wave absorption; rectangular waveguides; dielectric spacer; field intensity; head-equivalent liquid; output voltage; rectangular waveguide; reference antenna; reference dipole antenna; specific absorption rate-probe calibration system; tissue-equivalent liquid; Antenna measurements; Attenuation; Calibration; Dipole antennas; Liquid waveguides; Probes; Prototypes; Specific absorption rate; Transmitting antennas; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electromagnetic Compatibility and 19th International Zurich Symposium on Electromagnetic Compatibility, 2008. APEMC 2008. Asia-Pacific Symposium on
  • Conference_Location
    Singapore
  • Print_ISBN
    978-981-08-0629-3
  • Electronic_ISBN
    978-981-08-0629-3
  • Type

    conf

  • DOI
    10.1109/APEMC.2008.4559825
  • Filename
    4559825