DocumentCode
2074606
Title
Relaxofon: A neuromuscular blockade monitor for patients under general anesthesia
Author
Wehbe, Mohamad ; Mathieu, P.A. ; Hemmerling, Thomas M.
Author_Institution
Div. of Exp. Surg., McGill Univ., Montreal, QC, Canada
fYear
2012
fDate
Aug. 28 2012-Sept. 1 2012
Firstpage
150
Lastpage
153
Abstract
Anesthesia consists of three components: unconsciousness, analgesia and neuromuscular blockade (NMB). A specific drug is administered by the anesthesiologist to control these different components. In this paper we propose a new system for monitoring the neuromuscular blockade in anesthesized patients during surgery. Neuromuscular blockade drugs are used routinely by clinicians to induce muscle relaxation in patients. However, the use of these drugs has some risks, so an adequate monitoring of the effects of these drugs is essential. This paper describes the Relaxofon, a NMB monitoring device based on phonomyography. The Relaxofon is composed of a hardware subsystem that records muscle sounds using microphones and a special circuit to filter out the noise and amplify the signal, and a software subsystem that analyses the acquired signal. We tested the ability of the system to record phonomyographic signals from the adductor pollicis and the corrugator supercilii muscles. We then performed the Bland-Altman test to compare the manual Train-of-Four ratio (a measure of the depth of muscle relaxation) calculation against the one performed by the Relaxofon. Finally, we calculated the Pearson correlation coefficient to measure the linear dependence between the two methods. Automatic Train-of-Four ratio calculations using this system showed very good agreement with manual calculations. Results from this work may ultimately lead to integration of NMB monitoring to an automated closed-loop anesthesia system.
Keywords
acoustic signal detection; acoustic signal processing; biomedical equipment; drugs; filtering theory; medical signal detection; medical signal processing; microphones; muscle; neurophysiology; patient monitoring; signal denoising; statistical testing; surgery; Bland-Altman test; NMB monitoring device; Pearson correlation coefficient; Relaxofon; acquired signal analysis; adductor pollicis; analgesia; anesthesiologist; automated closed-loop anesthesia system; corrugator supercilii muscles; general anesthesia patients; hardware subsystem; manual train-of-four ratio; microphones; muscle relaxation; muscle sounds; neuromuscular blockade drugs; neuromuscular blockade monitoring; noise filtering; phonomyographic signals; signal amplification; software subsystem; surgery; unconsciousness; Anesthesia; Biomedical measurements; Manuals; Microphones; Monitoring; Neuromuscular; Anesthesia, General; Female; Humans; Male; Monitoring, Physiologic; Myography; Neuromuscular Blockade;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE
Conference_Location
San Diego, CA
ISSN
1557-170X
Print_ISBN
978-1-4244-4119-8
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/EMBC.2012.6345893
Filename
6345893
Link To Document