• DocumentCode
    2079688
  • Title

    Method of locating ultrasound-powered nerve stimulators

  • Author

    Gulick, Daniel W. ; Towe, Bruce C.

  • Author_Institution
    Sch. of Health & Biol. Syst. Eng., Arizona State Univ., Tempe, AZ, USA
  • fYear
    2012
  • fDate
    Aug. 28 2012-Sept. 1 2012
  • Firstpage
    887
  • Lastpage
    890
  • Abstract
    We have previously shown a small simple ultrasound-powered nerve stimulator. The piezoelectric implant receives power from an external driving ultrasound transducer. Focusing the ultrasound beam improves power transfer efficiency, but the implant location must be known to aim the focus. We show that currents driven by the stimulator might be detectable on the skin. By scanning the ultrasound focus and measuring the electrical response, we form an image of the implant location. This could give a feedback signal for aiming the beam, and allow multichannel addressing of several stimulators with no added circuitry in the implant.
  • Keywords
    neurophysiology; patient treatment; piezoelectric devices; prosthetics; ultrasonic focusing; ultrasonic transducers; piezoelectric implant; power transfer efficiency; ultrasound beam focusing; ultrasound-powered nerve stimulators; Electrodes; Implants; Phantoms; Receivers; Skin; Transducers; Ultrasonic imaging; Animals; Electric Stimulation Therapy; Electrodes, Implanted; Humans; Models, Biological; Peripheral Nerves; Sound;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-4119-8
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/EMBC.2012.6346074
  • Filename
    6346074