• DocumentCode
    2079831
  • Title

    Using analytical redundancy to increase safety of a synergistic manually guided instrument for craniotomy

  • Author

    Korff, Alexander ; Kellermann, C. ; Mueller, Matthias ; Kunze, S.C. ; Schmieder, Ken ; Radermacher, Klaus

  • Author_Institution
    Helmholtz Inst. for Biomed. Eng., Med. Eng., RWTH Aachen Univ., Aachen, Germany
  • fYear
    2012
  • fDate
    Aug. 28 2012-Sept. 1 2012
  • Firstpage
    912
  • Lastpage
    915
  • Abstract
    In this paper, two methods for bridging line of sight interruptions occurring during the use of the synergistically operated semiautomatic trepanation system (STS) are presented. In the STS, position information is acquired using an optical tracking system with the disadvantage of possible line-of-sight interruptions. Their compensation is crucial, as a real-time control system automatically adjusts the cutting depth of the instrument on the basis of position information and a-priori data. The surgeon is only responsible for guiding the instrument along the resection line. Hence, availability of position information is crucial for depth control, set point generation, and thus for patient safety. In favour of enhancing reliability of position and orientation acquisition, two approaches were developed which are intended to estimate the position during line of sight interruptions on the basis of a-priori system information and process parameters. To assure patient´s safety during this procedure, several parameters of the system (e.g. cutting radius, skull gradient) are used in order to estimate the possible cutting error while the redundant system is activated. These two algorithms and the online risk assessment were implemented, and afterwards evaluated. The evaluation was performed using a skull phantom, and yielded promising results.
  • Keywords
    biomedical equipment; brain; medical control systems; neurophysiology; optical tracking; phantoms; reliability; risk management; safety; surgery; STS; analytical redundancy; craniotomy; online risk assessment; optical tracking system; patient safety; real-time control system; sight interruptions; skull phantom; synergistic manually guided instrument; Adaptive optics; Instruments; Safety; Sensors; Surgery; Tracking; Trajectory; Craniotomy; Humans; Models, Biological; Safety;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-4119-8
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/EMBC.2012.6346080
  • Filename
    6346080