Title :
Novel bearingless centrifugal blood pump
Author :
Wan, S. ; Tseng, K.J. ; Chan, W.K.
Author_Institution :
Sch. of Electr. & Electron. Eng., Nanyang Technol. Univ., Singapore
Abstract :
A new directly-driven centrifugal blood pump has been designed and constructed. It is based on a special form of permanent magnet brushless DC motor whose rotor is combined with an impeller and encased in a sealed pump housing. To prevent potential thromboembolism and clotting, the rotor does not have any mechanical bearing. Instead, it is magnetically suspended in the pump housing. The multi-winding stator is located outside the pump housing and provides the rotor with rotation motion and also the magnetic suspension force. With the aid of three-dimensional finite element simulator, the motor has been simulated and analysed. This device may find application as a temporary blood pump or even as an implantable artificial heart.
Keywords :
artificial organs; brushless DC motors; finite element analysis; magnetic bearings; magnetic forces; pumps; rotors; stators; bearingless centrifugal blood pump; impeller; implantable artificial heart; magnetic suspension; magnetic suspension force; magnetically suspended bearing; multi-winding stator; permanent magnet brushless DC motor; rotation motion; rotor; sealed pump housing; temporary blood pump; three-dimensional finite element simulator; Analytical models; Blood; Brushless DC motors; Coagulation; Finite element methods; Impellers; Magnetic levitation; Mechanical bearings; Rotors; Stators;
Conference_Titel :
Power Electronics and Drive Systems, 2001. Proceedings., 2001 4th IEEE International Conference on
Print_ISBN :
0-7803-7233-6
DOI :
10.1109/PEDS.2001.975411