• DocumentCode
    2083298
  • Title

    Comparative evaluation of monocular augmented-reality display for surgical microscopes

  • Author

    Rodriguez Palma, Santiago ; Becker, Brian C. ; Lobes, L.A. ; Riviere, Cameron N.

  • Author_Institution
    Univ. of Valladolid, Valladolid, Spain
  • fYear
    2012
  • fDate
    Aug. 28 2012-Sept. 1 2012
  • Firstpage
    1409
  • Lastpage
    1412
  • Abstract
    Medical augmented reality has undergone much development recently. However, there is a lack of studies quantitatively comparing the different display options available. This paper compares the effects of different graphical overlay systems in a simple micromanipulation task with “soft” visual servoing. We compared positioning accuracy in a real-time visually-guided task using Micron, an active handheld tremor-canceling microsurgical instrument, using three different displays: 2D screen, 3D screen, and microscope with monocular image injection. Tested with novices and an experienced vitreoretinal surgeon, display of virtual cues in the microscope via an augmented reality injection system significantly decreased 3D error (p <; 0.05) compared to the 2D and 3D monitors when confounding factors such as magnification level were normalized.
  • Keywords
    augmented reality; computer displays; control engineering computing; medical computing; medical robotics; micromanipulators; microscopes; robot vision; screens (display); surgery; visual servoing; 2D screen; 3D screen; Micron; active handheld tremor-canceling microsurgical instrument; comparative evaluation; graphical overlay systems; medical augmented reality; micromanipulation task; monocular augmented-reality display; monocular image injection; real-time visually-guided task; robotic system; soft visual servoing; surgical microscopes; virtual cues display; vitreoretinal surgeon; Accuracy; Fixtures; Microscopy; Monitoring; Surgery; Three dimensional displays; Visualization; Equipment Design; Hand; Humans; Image Processing, Computer-Assisted; Microscopy; Signal Processing, Computer-Assisted; Surgery, Computer-Assisted; Tremor; User-Computer Interface;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-4119-8
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/EMBC.2012.6346203
  • Filename