DocumentCode :
2083298
Title :
Comparative evaluation of monocular augmented-reality display for surgical microscopes
Author :
Rodriguez Palma, Santiago ; Becker, Brian C. ; Lobes, L.A. ; Riviere, Cameron N.
Author_Institution :
Univ. of Valladolid, Valladolid, Spain
fYear :
2012
fDate :
Aug. 28 2012-Sept. 1 2012
Firstpage :
1409
Lastpage :
1412
Abstract :
Medical augmented reality has undergone much development recently. However, there is a lack of studies quantitatively comparing the different display options available. This paper compares the effects of different graphical overlay systems in a simple micromanipulation task with “soft” visual servoing. We compared positioning accuracy in a real-time visually-guided task using Micron, an active handheld tremor-canceling microsurgical instrument, using three different displays: 2D screen, 3D screen, and microscope with monocular image injection. Tested with novices and an experienced vitreoretinal surgeon, display of virtual cues in the microscope via an augmented reality injection system significantly decreased 3D error (p <; 0.05) compared to the 2D and 3D monitors when confounding factors such as magnification level were normalized.
Keywords :
augmented reality; computer displays; control engineering computing; medical computing; medical robotics; micromanipulators; microscopes; robot vision; screens (display); surgery; visual servoing; 2D screen; 3D screen; Micron; active handheld tremor-canceling microsurgical instrument; comparative evaluation; graphical overlay systems; medical augmented reality; micromanipulation task; monocular augmented-reality display; monocular image injection; real-time visually-guided task; robotic system; soft visual servoing; surgical microscopes; virtual cues display; vitreoretinal surgeon; Accuracy; Fixtures; Microscopy; Monitoring; Surgery; Three dimensional displays; Visualization; Equipment Design; Hand; Humans; Image Processing, Computer-Assisted; Microscopy; Signal Processing, Computer-Assisted; Surgery, Computer-Assisted; Tremor; User-Computer Interface;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE
Conference_Location :
San Diego, CA
ISSN :
1557-170X
Print_ISBN :
978-1-4244-4119-8
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/EMBC.2012.6346203
Filename :
6346203
Link To Document :
بازگشت