DocumentCode
2086798
Title
A Multi-Channel Generator at 433 MHz with Accurate Amplitude and Phase Control for Hyperthermia Application
Author
Martens, L. ; De Wagter, C.
Author_Institution
Laboratory of Electromagnetism and Acoustics, University of Gent, Sint-Pietersnieuwstraat 41, B-9000 Gent, Belgium
Volume
2
fYear
1991
fDate
9-12 Sept. 1991
Firstpage
1433
Lastpage
1437
Abstract
A new generator for hyperthermia application is presented with feedback control based on a Phased Locked Loop (PLL). The phase and amplitude can be controlled in a very accurate way since the phase and the amplitude are locked by direct feedback from the applicator. The phase and amplitude shifts in the amplifiers are compensated by the phase and amplitude controlling. The generator will be used in a multi-channel hyperthermiia system. Accurate phase and amplitude control at the applicator stage is a necessary condition for the evaluation of experiments with phased arrays. The accuracy of the phase and amplitude steering is proved by measurements with a network analyzer.
Keywords
Applicators; Detectors; Filters; Frequency conversion; Hyperthermia; Phase control; Phase detection; Phase locked loops; Phased arrays; Voltage-controlled oscillators;
fLanguage
English
Publisher
ieee
Conference_Titel
Microwave Conference, 1991. 21st European
Conference_Location
Stuttgart, Germany
Type
conf
DOI
10.1109/EUMA.1991.336546