Abstract :
According to the morphologic character of cells, a simple way based on the circular aperture diffraction is put forward in the paper. Cells in different states, such as normal living cells, apoptosis cells, necrosis cells, can be differentiated easily and the quantities and the quantities ratio of cells in different states can be got in this way. The diameter of sample cell can be measured accurately by the circular aperture diffraction combined with CCD. The frame diagram of measuring the size of cells based on the circular diffraction and CCD is presented, as well as the Principle of method in the paper. Using the method presented in the paper, the apoptosis and necrosis situations of leukaemia tumour cell HL60 were measured. The experiment results showed that the results of quantities and the quantities ratios of HL60 cells in different states such as apoptosis and necrosis measured by the method presented in the paper were consistent with the results measured by the other accepted method.
Keywords :
biomedical optical imaging; blood; cellular biophysics; light diffraction; tumours; CCD; apoptosis cells; cell diameter; circular aperture diffraction; leukaemia tumour cell HL60; necrosis cells; normal living cells; optical manner; tumor cell identification; Apertures; Biomedical measurements; Biomedical optical imaging; Cells (biology); Charge coupled devices; Electron microscopy; Neoplasms; Optical diffraction; Optical microscopy; Tumors; CCD; cell; diffraction;
Conference_Titel :
Complex Medical Engineering, 2007. CME 2007. IEEE/ICME International Conference on
Conference_Location :
Beijing
Print_ISBN :
978-1-4244-1077-4
Electronic_ISBN :
978-1-4244-1078-1
DOI :
10.1109/ICCME.2007.4381909