Title :
A digital instrument for venous muscle pump test
Author :
Bhooma, G. ; Kokila, S. ; Jayanthi, K. ; Kumar, Varadarajan Jagadeesh
Author_Institution :
Dept. of Electron. & Commun. Eng., Pondicherry Eng. Coll., Pondicherry, India
fYear :
2011
fDate :
10-12 May 2011
Firstpage :
1
Lastpage :
4
Abstract :
Design and development of a prototype unit to measure venous refilling time (VRT) and muscle pump capacity (MPC) during a venous muscle pump test (VMPT) is presented here. Instead of normalizing the photo-plethysmo-graph (PPG), as is common in the VMPT instruments in vogue, a suitable signal processing technique that extracts a signal proportional only to the blood volume changes underneath the PPG sensor is employed. Apart from the measurement of VRT and MPC, the prototype unit also provides a measure on the pulse rate.
Keywords :
biomedical equipment; blood pressure measurement; blood vessels; muscle; photoplethysmography; PPG sensor; blood volume changes; digital instrument; muscle pump capacity; photoplethysmo-graph; prototype unit; pulse rate measurement; signal processing technique; signal proportionality; venous muscle pump test; venous refilling time; Blood; Detectors; Muscles; Prototypes; Pulse measurements; Software; Veins; Biomedical measurements; Biomedical transducers; Optical sensors; Photoplethysmography; Venous refilling time; microcontroller; muscle pump capacity;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Instrumentation and Measurement Technology Conference (I2MTC), 2011 IEEE
Conference_Location :
Binjiang
ISSN :
1091-5281
Print_ISBN :
978-1-4244-7933-7
Type :
conf
DOI :
10.1109/IMTC.2011.5944063