DocumentCode :
2098881
Title :
Single transmit imaging
Author :
Liu, D-L Donald ; Behren, Pat Von ; Kim, Jin
Author_Institution :
Ultrasound Group, Siemens Med. Syst., Issaquah, WA, USA
Volume :
2
fYear :
1999
fDate :
1999
Firstpage :
1275
Abstract :
Single transmit imaging, formed by scanning the receive beam around a single transmit beam is used to study the characteristics of fundamental and harmonic beams and to evaluate the performance of time-shift compensation as a means of phase aberration correction. The results indicate that the harmonic beam has a greater depth of field compared to the normal beam, and that time-shift compensation can restore the beam pattern using random scattering data under ideal aberration conditions. Results of time-shift compensation using random scattering data also showed significant improvement in image quality on a special phase aberration phantom that contains an aberration layer of extended thickness
Keywords :
biomedical ultrasonics; ultrasonic scattering; beam pattern restoration; extended thickness aberration layer; ideal aberration conditions; image quality improvement; medical diagnostic imaging; random scattering data; special phase aberration phantom; time-shift compensation; time-shift compensation performance evaluation; Acoustic beams; Acoustic scattering; Biomedical imaging; Convolution; Data acquisition; Frequency; Image quality; Image restoration; Imaging phantoms; Ultrasonic imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 1999. Proceedings. 1999 IEEE
Conference_Location :
Caesars Tahoe, NV
ISSN :
1051-0117
Print_ISBN :
0-7803-5722-1
Type :
conf
DOI :
10.1109/ULTSYM.1999.849229
Filename :
849229
Link To Document :
بازگشت