DocumentCode
2099242
Title
Hermetic electronic packaging of an implantable brain-machine-interface with transcutaneous optical data communication
Author
Schuettler, Martin ; Kohler, Fabian ; Ordonez, Juan S. ; Stieglitz, T.
Author_Institution
Dept. of Microsyst. Eng., Univ. of Freiburg, Freiburg, Germany
fYear
2012
fDate
Aug. 28 2012-Sept. 1 2012
Firstpage
3886
Lastpage
3889
Abstract
Future brain-computer-interfaces (BCIs) for severely impaired patients are implanted to electrically contact the brain tissue. Avoiding percutaneous cables requires amplifier and telemetry electronics to be implanted too. We developed a hermetic package that protects the electronic circuitry of a BCI from body moisture while permitting infrared communication through the package wall made from alumina ceramic. The ceramic package is casted in medical grade silicone adhesive, for which we identified MED2-4013 as a promising candidate.
Keywords
adhesives; biological tissues; biomedical electronics; biomedical telemetry; brain; brain-computer interfaces; ceramic packaging; handicapped aids; optical communication; prosthetics; telemedicine; MED2-4013; alumina ceramic; amplifier; brain tissue; brain-computer-interfaces; electrical contact; electronic circuitry; hermetic electronic packaging; implantable brain-machine-interface; infrared communication; medical grade silicone adhesive; package wall; telemetry electronics; transcutaneous optical data communication; Adhesives; Implants; Packaging; Rubber; Substrates; Wires; Adhesiveness; Brain; Communication; Dimethylpolysiloxanes; Electrodes, Implanted; Electronics, Medical; Humans; Infrared Rays; Product Packaging; Silicones; Skin; User-Computer Interface;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE
Conference_Location
San Diego, CA
ISSN
1557-170X
Print_ISBN
978-1-4244-4119-8
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/EMBC.2012.6346816
Filename
6346816
Link To Document