DocumentCode
2111839
Title
Electrodes´ erosion in a liquid medium [US lithotriptor]
Author
Allagui, Hatem
Author_Institution
L.S.E, Ecole Nationale d´Ingenieurs de Tunis
Volume
4
fYear
2002
fDate
2002
Firstpage
1289
Abstract
We seek by this work to tackle the problems of electrodes´ durability of a lithotriptor, which is intended to break a renal calculus. These electrodes functioning in a liquid medium are made up of a composite material. However, in spite of their high prices, these electrodes bear only 3000 electric discharges corresponding to a net time utilization of approximately one hour. It is thus a question of looking for reducing the cost of lithotherapeutics exploitation by replacing the used composite material by a homogenous material which is available in the local market even if it is less resistant to erosion than the first.
Keywords
biomedical electrodes; biomedical ultrasonics; kidney; radiation therapy; wear resistance; 1 hr; composite material; electric arc current; electric discharges; electrode erosion; electrodes durability; homogenous material; lithotherapeutics exploitation costs reduction; lithotriptor; renal calculus breaking;