Title :
Commissioning results of the LLUMC beam switchyard and gantry
Author_Institution :
Sci. Applications Int. Corp., San Diego, CA, USA
Abstract :
Results from the commissioning of the LLUMC (Loma Linda University Medical Center) beam switchyard and gantry are described. Measurements of the beam profile at numerous locations are compared with model predictions for three energies. On the basis of the beam profile and magnetic measurements, the extracted beam emittance and Twiss parameters are determined.<>
Keywords :
beam handling equipment; particle beam diagnostics; LLUMC beam switchyard; Loma Linda University Medical Center; Twiss parameters; beam gantry; beam profile; extracted beam emittance; magnetic measurements; Biomedical optical imaging; Energy measurement; Extraterrestrial measurements; Magnets; Medical treatment; Optical beams; Particle beams; Predictive models; Proton accelerators; Stimulated emission;
Conference_Titel :
Particle Accelerator Conference, 1991. Accelerator Science and Technology., Conference Record of the 1991 IEEE
Conference_Location :
San Francisco, CA, USA
Print_ISBN :
0-7803-0135-8
DOI :
10.1109/PAC.1991.164379