DocumentCode
2119194
Title
Real-time tagless monitoring of cell viability using patch-clamp microdevices
Author
Pathak, Pushparaj ; Zhao, Hong ; Gong, Zhongcheng ; Nie, Fang ; Zhang, Tianhua ; Wong, Stephen ; Que, Long
Author_Institution
Instn. for Micromanufacruring, Louisiana Tech Univ., Ruston, LA, USA
fYear
2010
fDate
1-4 Nov. 2010
Firstpage
2482
Lastpage
2485
Abstract
Real-time tagless monitoring of cell viability using patch-clamp microdevices is reported and validated by using fluorescence imaging techniques for the first time. Specifically, human breast cancer cells MDA-MB-231 have been used for these studies. Systematic experiments on these cells found that the impedance change of captured cells by the micro-electrodes is a valid indicator of cell´s living or death, which might provide an efficient way to monitor the drug effects on cells, thereby benefiting high throughput drug screen and discovery process.
Keywords
bioMEMS; biological organs; biomedical optical imaging; cancer; cellular biophysics; drugs; fluorescence; gynaecology; microelectrodes; cell viability; drug effects; fluorescence imaging; high throughput drug screen; human breast cancer cells; impedance change; microelectrodes; patch-clamp microdevices; real-time tagless monitoring;
fLanguage
English
Publisher
ieee
Conference_Titel
Sensors, 2010 IEEE
Conference_Location
Kona, HI
ISSN
1930-0395
Print_ISBN
978-1-4244-8170-5
Electronic_ISBN
1930-0395
Type
conf
DOI
10.1109/ICSENS.2010.5690086
Filename
5690086
Link To Document